Agreement between artificial intelligence-assisted and clinician-derived proximal femoral anatomical measurements on computed tomography
Accurate preoperative planning for total hip arthroplasty (THA) requires precise identification of proximal femoral anatomical landmarks, which remains observer-dependent. Although machine learning software can automate landmark detection, comparisons with clinician variability are scarce. This study evaluated whether automated landmark extraction on computed tomography (CT) scans falls within the range of human interobserver variability. Forty-three CT scans from patients undergoing preoperative planning for primary THA were analyzed. Three-dimensional femoral models were reconstructed in the CHIDOS software, which automatically segmented bone structures and extracted anatomical landmarks. Automated measurements were compared with manual measurements independently performed by two orthopedic consultants. Agreement was assessed using Euclidean deviations, intraclass correlation coefficients (ICC), Bland–Altman analysis, and paired Wilcoxon signed-rank tests, supplemented by an exploratory post-hoc equivalence analysis using two one-sided t-tests (TOST). Mean Euclidean deviations for landmark localization ranged from 2.79 ± 1.47 mm to 5.01 ± 1.73 mm, and angular deviations ranged from 1.33 ± 0.97° to 2.38 ± 1.75°. Agreement for angular measurements was good to excellent (ICC: 0.89–0.98). Artificial intelligence (AI)–clinician deviations were lower for femoral anteversion (p < 0.001) and neck-shaft angle (p = 0.004) relative to Clinician 2, higher for femoral head center (p < 0.001) and lesser trochanter (p = 0.008), and lower for femoral neck isthmus (p = 0.006) relative to Clinician 1. The 95% confidence intervals of the mean deviation relative to interobserver variability lay within literature-reported margins (±2° angular, ±3 mm linear) for all five endpoints. Post-hoc TOST analysis supported this observation (p < 0.001 for all endpoints), but did not establish clinical equivalence or non-inferiority. These preliminary findings support further evaluation of AI-assisted landmark extraction as a reproducible first-pass step in CT-based THA planning workflows; effects on implant selection, operative performance, and clinical outcomes remain to be demonstrated.

- Guo L, Naghavi SA, Wang Z, et al. On the design evolution of hip implants: a review. Mater Des. 2022;216:110552. doi: 10.1016/j.matdes.2022.110552
- Huppertz A, Radmer S, Wagner M, Roessler T, Hamm B, Sparmann M. Computed tomography for preoperative planning in total hip arthroplasty: what radiologists need to know. Skeletal Radiol. 2014;43(8):1041-1051. doi: 10.1007/s00256-014-1853-2
- Parisi FR, Zampogna B, Zampoli A, et al. Planning accuracy and stem offset assessment in digital two-dimensional versus three-dimensional planning in cementless hip arthroplasty: a systematic review and meta-analysis. J Clin Med. 2024;13(21):6566. doi: 10.3390/jcm13216566
- Moralidou M, Di Laura A, Henckel J, Hothi H, Hart AJ. Accuracy of a three-dimensional (3D)-printed patient-specific (PS) femoral osteotomy guide: a computed tomography (CT) study. Bioengineering. 2022;9(11):667. doi: 10.3390/bioengineering9110667
- Schmid J, Chênes C, Chagué S, et al. MyHip: supporting planning and surgical guidance for a better total hip arthroplasty. Int J Comput Assist Radiol Surg. 2015;10(10):1547-1556. doi: 10.1007/s11548-015-1177-2
- Lattanzi R, Viceconti M, Zannoni C, Quadrani P, Toni A. Hip-Op: an innovative software to plan total hip replacement surgery. Med Inform Internet Med. 2002;27(2):71-83. doi: 10.1080/14639230210150346
- Eskandar K. The role of artificial intelligence in orthopedic surgery: current applications and future perspectives—a systematic review of the literature. Rev Esp Cir Ortop Traumatol. 2026;70(3):248-257. doi: 10.1016/j.recot.2025.09.005
- Khaje Mozafari J, Moshtaghioon SA, Mahdavi SM, Ghaznavi A, Behjat M, Yeganeh A. The role of artificial intelligence in preoperative planning for total hip arthroplasty: a systematic review. Front Artif Intell. 2024;7:1417729. doi: 10.3389/frai.2024.1417729
- Wang D. Clinical validation of the segmentation algorithms underlying an automatic surgery planning system for total hip arthroplasty. Conference abstract. European Congress of Radiology; 2025. doi: 10.26044/ECR2025/C-16661
- Wang H, Zhang X, Li S, et al. A 3D multi-task network for the automatic segmentation of CT images featuring hip osteoarthritis. Biomed Phys Eng Express. 2025;11(5):055042. doi: 10.1088/2057-1976/ae0593
- Spinciu IT, Dușa S, Petean CE, Sultana AE, Cristea DD, Liciu E. Automatic multi-class segmentation and landmark-based osteotomy planning for hip arthroplasty. In: Proceedings of the 2025 IEEE 21st International Conference on Intelligent Computer Communication and Processing (ICCP). IEEE; 2025:1-9. doi: 10.1109/ICCP68926.2025.11427145
- Wan J, Liu L, Wang H, et al. UNSCT-HRNet: modeling anatomical uncertainty for landmark detection in total hip arthroplasty. Preprint. arXiv; 2024. doi: 10.48550/arXiv.2411.08488
- Zhai H, Chen Z, Li L, et al. Two-stage multi-task deep learning framework for simultaneous pelvic bone segmentation and landmark detection from CT images. Int J Comput Assist Radiol Surg. 2024;19(1):97-108. doi: 10.1007/s11548-023-02976-1
- Li L, Ye Q, Hu T, Fang D, Wan R, Shi Y. Fully automated measurement of femoral neck axis line in anatomical coordinate system using CT images. Med Biol Eng Comput. 2024;62(6):1837-1849. doi: 10.1007/s11517-024-03037-8
- Schmaranzer F, Movahhedi M, Singh M, et al. Computed tomography-based automated 3D measurement of femoral version: validation against standard 2D measurements in symptomatic patients. J Orthop Res. 2024;42(10):2237-2248. doi: 10.1002/jor.25865
- Xiao H, Yibulayimu S, Zhao C, et al. Rethinking femoral neck anteversion assessment: a novel automated 3D CT method compared to traditional manual techniques. BMC Musculoskelet Disord. 2025;26(1):466. doi: 10.1186/s12891-025-08697-7
- Chen X, Liu X, Wang Y, et al. Development and validation of an artificial intelligence preoperative planning system for total hip arthroplasty. Front Med. 2022;9:841202. doi: 10.3389/fmed.2022.841202
- Wu L, Zhao X, Lu Z, Yang Y, Ma L, Li P. Accuracy analysis of artificial intelligence-assisted three-dimensional preoperative planning in total hip replacement. Jt Dis Relat Surg. 2023;34(3):537-547. doi: 10.52312/jdrs.2023.1059
- Chan PYW, Baker CE, Suh Y, Moyer D, Martin JR. Development of a deep learning model for automating implant position in total hip arthroplasty. J Arthroplasty. 2025;40(8):2092-2100. doi: 10.1016/j.arth.2025.01.032
- Xie H, Yi J, Huang Y, et al. Application and evaluation of artificial intelligence 3D preoperative planning software in developmental dysplasia of the hip. J Orthop Surg Res. 2024;19(1):176. doi: 10.1186/s13018-024-04588-0
- Clement ND, Clement R, Clement A. Predicting functional outcomes of total hip arthroplasty using machine learning: a systematic review. J Clin Med. 2024;13(2):603. doi: 10.3390/jcm13020603
- Rostamian R, Shariat Panahi M, Karimpour M, Kashani HG, Abi A. A deep learning-based multi-view approach to automatic 3D landmarking and deformity assessment of lower limb. Sci Rep. 2025;15(1):534. doi: 10.1038/s41598-024-84387-z
- Berger L, Brößner P, Ehreiser S, Tokunaga K, Okamoto M, Radermacher K. Validation and comparison of three different methods for automated identification of distal femoral landmarks in 3D. Biomed Tech. 2025;70(5):425-431. doi: 10.1515/bmt-2025-0026
- Jang SJ, Kunze KN, Vigdorchik JM, Jerabek SA, Mayman DJ, Sculco PK. John Charnley Award: deep learning prediction of hip joint center on standard pelvis radiographs. J Arthroplasty. 2022;37(7):S400-S407.e1. doi: 10.1016/j.arth.2022.03.033
- Vidhani FR, Woo JJ, Zhang YB, Olsen RJ, Ramkumar PN. Automating linear and angular measurements for the hip and knee after computed tomography: validation of a three-stage deep learning and computer vision-based pipeline for pathoanatomic assessment. Arthroplasty Today. 2024;27:101394. doi: 10.1016/j.artd.2024.101394
- Edgar HJH, Daneshvari Berry S, Moes E, Adolphi NL, Bridges P, Nolte KB. New Mexico Decedent Image Database. Dataset. Published online 2020. doi: 10.25827/5s8c-n515
- Wolf I, Vetter M, Wegner I, et al. The Medical Imaging Interaction Toolkit. Med Image Anal. 2005;9(6):594-604. doi: 10.1016/j.media.2005.04.005
- Lee YS, Oh SH, Seon JK, Song EK, Yoon TR. 3D femoral neck anteversion measurements based on the posterior femoral plane in ORTHODOC® system. Med Biol Eng Comput. 2006;44(10):895-906. doi: 10.1007/s11517-006-0104-7